UPV Digital RepositoryUPV-DRUniversity of the Philippines Visayas
 

Determination of instrument detection limit and method detection limit of Varian 55B flame atomic absorption spectrometer for cadmium, chromium, copper, iron, lead, manganese and zinc using sips-10 and manual sample introduction methods

dc.contributor.advisorTaberna, Hilario, Jr.
dc.contributor.authorCalawigan, Roland Joseph C.
dc.contributor.chairCainglet, Rica P.
dc.contributor.committeememberGamarcha, Leandro
dc.contributor.committeememberPrimavera, Karlo H.
dc.date.accessioned2026-09-22T07:39:12Z
dc.date.issued2009-04
dc.description.abstractDetermination of detection limits is an essential measure of performance for a quantitative laboratory, method or analyst. In this study, the Instrument Detection Limit (IDL) and Method Detection Limit (MDL) of F-AAS Varian 55B for cadmium, chromium, copper, iron, lead, manganese and zinc were determined using the SIPS-10 and manual sample introduction methods. For IDL determinations, EPA’s SW-846 method (Test Methods for Evaluating Solid Waste and Physical/Chemical Methods) was employed for both sample introduction methods where deionized water was analyzed for twenty times for three non-consecutive days. MDL determinations were done based on the definition given by the United States Code of Federal Regulations (40 CFR part 136, Appendix B - Definition and Procedure for the Determination of the Method Detection Limit). In this procedure, MDLs were estimated by calculating the standard deviation from the analysis of a standard solution containing the analyte with a concentration determined from the estimated detection limit, which was also analyzed for twenty times. Results showed that for the determination of IDL, all of the elements have lower values (Cd 0.000887 mg/L, Cr 0.00435 mg/L, Cu 0.000128 mg/L, Fe 0.000816 mg/L, Pb 0.000742 mg/L, Mn 0.000985 mg/L, Zn 0.000585 mg/L) for the manual method than the SIPS-10 method (Cd 0.00399 mg/L, Cr 0.0150 mg/L, Cu 0.00222 mg/L, Fe 0.00913 mg/L, Pb 0.0217 mg/L, Mn 0.00132 mg/L, Zn 0.00445 mg/L). For MDL, the results showed variations for both the two sample introduction methods. The MDL values for cadmium, chromium, copper, iron, lead, manganese and zinc are 0.0374 mg/L, 0.0235 mg/L, 0.00203 mg/L, 0.0197 mg/L, 0.0805 mg/L, 0.00325 mg/L and 0.00686 mg/L in the manual method, respectively and 0.0101 mg/L, 0.0429 mg/L, 0.00280 mg/L, 0.0186 mg/L, 0.19 7 mg/L, 0.00312 mg/L and 0.0103 mg/L in the SIPS-10 method, respectively. Both methods have comparable precision for copper, iron, manganese and zinc determinations while SIPS-10 method is more precise for cadmium, chromium, and lead analysis.
dc.format.extentix, 48 p.
dc.identifier.citationCalawigan, R. J. C. (2009). <i>Determination of instrument detection limit and method detection limit of Varian 55B flame atomic absorption spectrometer for cadmium, chromium, copper, iron, lead, manganese and zinc using sips-10 and manual sample introduction methods</i> [Undergraduate thesis, University of the Philippines Visayas]. UPV Institutional Repository. https://hdl.handle.net/20.500.14583/850
dc.identifier.urihttps://hdl.handle.net/20.500.14583/850
dc.publisherDepartment of Chemistry, College of Arts and Sciences, University of the Philippines Visayas
dc.subjectFlame atomic absorption spectrometry
dc.subjectInstrument detection limit
dc.subjectMethod detection limit
dc.subjectAnalytical methods
dc.subjectTrace metals
dc.subject.agrovocatomic absorption spectrometry
dc.subject.agrovocdetection
dc.subject.agrovocanalytical methods
dc.subject.agrovoctrace elements
dc.subject.agrovoclead
dc.subject.lccLG 993.5 2009 C5 C36
dc.subject.sdgSDG 9 - Industry, innovation and infrastructure
dc.subject.sdgSDG 12 - Responsible consumption and production
dc.subject.sdgSDG 3 - Good health and well-being
dc.titleDetermination of instrument detection limit and method detection limit of Varian 55B flame atomic absorption spectrometer for cadmium, chromium, copper, iron, lead, manganese and zinc using sips-10 and manual sample introduction methods
dc.typeThesis
thesis.degree.departmentCollege of Arts and Sciences
thesis.degree.grantorUniversity of the Philippines Visayas
thesis.degree.levelUndergraduate
thesis.degree.nameBachelor of Science in Chemistry

Files

Original bundle

Now showing 1 - 1 of 1
Thumbnail Image
Name:
UPV-M-BSC-UT-2009-CalawiganRJC.pdf
Size:
74.33 KB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: